Phadia 100 - User Manual
Error Handling
12-3501-20/12
Page 12.23
12
Follow instructions in Action List in order to identify or eliminate the errors.
Check/remove if the movement of the Sample Carousel has been obstructed.
Make sure that nothing is blocking the movement by rotating the Sample
Carousel. Switch the Phadia 100 off, wait 10 seconds and switch on again. Clean
under the sample carousel if needed.
Rerun the results that have not been reported. Check that sample volume is
enough including the sample tube dead volume, otherwise new patient sample
is needed. Use
1.1.Enter requests
and put in the requests that have not been
reported.
If the error remains contact your local Phadia representative.
FATAL #16.07 WASH RAKE MOTOR CAN'T FIND FLAG
What can cause the problem:
The Wash Rake motor does not seem to work. Make sure that nothing is blocking
the up and down movement of the Wash Rake.
Corrective actions:
For Phadia 100 Ver.3.0, use the menu
6.13.Shutdown
and follow the instruc-
tions displayed to safely shut down the instrument. Switch Phadia 100 off, wait
10 seconds and switch on again.
Follow instructions in Action List in order to identify or eliminate the errors.
Rerun the results that have not been reported. Check that sample volume is
enough including the sample tube dead volume, otherwise new patient sample
is needed. Use
1.1.Enter requests
and put in the requests that have not been
reported.
If the error remains contact your local Phadia representative.
FATAL #16.08 CAP TRANSFER MOTOR CAN'T FIND FLAG
The ImmunoCAP Transfer motor does not seem to work.
Contact your local Phadia representative.
#17.XX OUT OF RANGE
What can cause the problem:
One of the motors is not correctly calibrated.
#17.01-17.08
Sub code see Stepper Motor table
Corrective actions:
Follow instructions in Action List in order to identify or eliminate the errors.
If the error remains contact your local Phadia representative.
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...