Phadia 100 - User Manual
Maintenance
12-3501-20/12
Page 11.5
2
5
6
11
available, leave to dry upside down, preferably not on paper (paper dust can be
fluoroscent).
Wipe off the Processing Chamber with a tissue. If a lot of water remains in the
Processing Chamber, highten the instrument by unscrewing the feet a few
millimeters on the printer side. It is important to keep the Processing Chamber
clean and dry.
Wipe off the underside of the Processing Lid with purified water in order to secure
the ImmunoCAP/ELiA Well transfer.
Install the Elution Wells Disc and ImmunoCAP Holder Disc. Make sure not to
exchange the Elution Wells Disc between instruments.
NOTE!
Before closing the Processing Lid, check that the wash rake is well above
the ImmunoCAP Carousel, otherwise it might break something inside. The wash
rake can be moved manually.
WARNING! Phadia 100 processing chamber is
exposed to human body fluids, which might be
infected. When servicing and maintaining Phadia
100, take precautions to avoid direct contact
with parts that have been in contact with
samples. Wear protective gloves!
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...