Phadia 100 - User Manual
Error Handling
12-3501-20/12
Page 12.57
12
been recapped. Check that the tubings to the bottle filters are connected properly.
See Chapter 11, Semiannual Maintenance.
7
77
77 Check Wash/Rinse: Consumption, Tubings:Connections
and that they are not twisted
Make sure that the tubing from the Washing and Rinse Solution bottles are not
twisted and that the tubings are properly connected to the bottles. To avoid
twisted tubes connect tubings after Washing and Rinse Solution bottles have
been recapped.
NOTE!
When running ImmunoCAP Specific IgA or ImmunoCAP Specific IgG
with a full sample carousel, 1 l of Washing Solution is not enough. Fill up the
bottle with Washing Solution.
Check how much Washing and Rinse Solution is left. If the consumption is
unreasonably low, contact your local Phadia representative.
8
88
88 Check/clean the Pipette tip - See User Manual,
Index: Pipette, clean
See Processing Errors #27.XX and #28.XX in this chapter.
9
99
99 Check/Remove if the movement of the ImmunoCAP Car-
ousel has been obstructed
Check the processing chamber for loose ImmunoCAP/EliA Wells, look under the
Elution Wells Disc check also under the Wash Rake. Clean the processing
chamber. If problem persist call service.
10
10
10
10
10 Check/remove misplaced ImmunoCAP/Wells in the
ImmunoCAP Carousel
Open processing lid carefully to see that all ImmunoCAP/EliA Wells are in the
inner position of the ImmunoCAP holder disc. If the positioning is not correct for
one ImmunoCAP/EliA Well but it is in the right line, correct the positioning after
checking that the identity is correct according to the Distribution List. Put in new
ImmunoCAP/EliA Wells manually if the identity is not possible to check or
several ImmunoCAP/EliA Wells are loose.
11
11
11
11
11 The Cuvette is probably broken
Check visually if the Cuvette is broken. Is the Cuvette broken, clean up the
Processing Chamber. Make sure there is no glass in the process chamber, check
Elution Wells Disc, lift up and look underneath also. Look for misplaced
ImmunoCAP/EliA Well on Elution Wells Disc, under Elution Wells Disc and
under Wash Rake. If ImmunoCAP/EliA Well is found, that is probably the
cause. If
ImmunoCAP not dispensed
and
Check ImmunoCAP Dispense
have
occurred prior to this error situation. That can be an indication of future
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...