Phadia 100 - User Manual
Assay Run Quality Control
12-3501-20/12
Page 6.11
6
4.4.2.Accept Calibrators
[sIgE]
Acceptance: OK?
4.4.2.Accept Calibrators
[sIgE]
sIgE calibrators: 090405 11:13 15?
Warnings that may prevent entering
NO CALIBRATORS EXIST
There are no stored calibrators for the selected method.
4.
Select the calibrators you
want by pressing
no
(or
no
and
shift
)
until the ones you want
appear, then press
yes
.
5.
Press
no
until the status
you want appears and then
press
yes
. The status will be
set.
Edit Calibrator Point
This is a laboratory password protected function that means that you have to
log in with the laboratory password. You can change the response for a
calibrator point.
1.
In the Edit Calibrators
mode press
3
.
2.
Press
yes
.
Warnings that may prevent entering
NO CALIBRATORS EXIST
There are no stored calibrators for the selected method.
3.
Select the calibrators you
want by pressing
no
until the
ones you want turn up, then
press
yes
.
The calibrators are identified by method, date and time for process and
calibration code.
4.4.Edit Calibrators
[sIgE]
3.Edit Calibrator Point?
4.4.3.Edit Calibrator Point
[sIgE]
sIgE calibrators: 090405 11:13 15?
4.4.Edit Calibrators
[sIgE]
1.Set Active Calibrators?
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...