Phadia 100 - User Manual
Instrument Specification and Installation
12-3501-20/12
Page 1.19
1
1
NOTE! Phadia 100 must
only be connected to a
grounded power outlet
(protective ground).
8.
Take out the Sample
Carousel from the acces-
sory box. Place it on the
left side in front of the
bottles. The vial holders
and the sample holders
are in their positions.
9.
Choose the proper
mains kit to comply with
your mains voltage. You
find the mains kit and the
fuse kit in the accessory box.
10.
Select the correct mains kit.
Discard the unwanted kit imme-
diately. Remove the warning la-
bel covering the fuse/voltage se-
lector unit on the rear panel of
the instrument.
Open the fuse/voltage selector
cover using a screwdriver. Insert
the screwdriver at the top centre
and apply pressure.
11.
Turn the power on in order to open the processing chamber. The chamber
is opened by pressing the
lid open
key.
Place the enclosed fuses in the
fuse holders and install the fuse
holders in both the left and right
hand position.
Remove the voltage selector
switch and select the correct
mains voltage by turning the
voltage selector. Place the
switch in the instrument. Close
the cover and make sure that
the selected voltage is visible
through the window. Connect
the main supply using the en-
closed cord.
6.
Remove the tape from
the Pipette Arm. Gently lift
the arm straight up and move
it to the right, away from the
Sample Carousel.
7.
Remove the tape from the Sample Carousel Cover and the tape
around the Rinse Solution, Washing Solution and Waste bottles.
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...