background image

MODEL 61302V

BAROMETRIC PRESSURE SENSOR

INSTRUCTIONS

R.M. YOUNG COMPANY

2801 AERO PARK DRIVE, TRAVERSE CITY, MICHIGAN  49686, USA

TEL: (231) 946-3980    FAX: (231) 946-4772

PN: 61302V-90

 REV: I111215

1.0 SPECIFICATIONS

Pressure 

500 to 1100 hPa standard range

Digital Accuracy* 

0.2 hPa (25°C)

 

0.3 hPa (-40°C to +60°C)

Analog Accuracy** 

0.05% of analog pressure range

Analog Temperature

   Dependence** 

0.0017% of analog pressure 

range per °C (25°C reference)

Long Term Stability:  0.2% FS per year

  *Defined as ±1 standard deviation from NIST-traceable 

pressure reference in clean, dry air.  Includes non-

linearity,  hysteresis,  repeatability,  and  calibration 

uncertainty.

 **Defined as ±1 standard deviation from ideal analog 

output.  Total analog output accuracy is the root sum 

square of digital accuracy, analog accuracy, and 

analog temperature dependence.

 

Output Rate 

1.8 Hz (max) to 1 per minute

 

Voltage Output 

0 to 5000 mV (standard) 

0 to 2500 mV 

12-bit resolution (1 in 4000) 

Selectable pressure range: 

500 to 1100 hPa (standard)

Serial Output 

Full duplex RS-232 

1200 to 38400 baud 

Continuous ASCII text 

Polled ASCII text 

NMEA 

0.01 hPa resolution

Supply Voltage 

7 to 30 VDC 

2.8 mA with Vout (standard) 

1.4 uA shutdown (Trig=0V)  

7 mA with serial I/O

Case 

Fiber-reinforced thermoplastic

Weight 

44 g (1.5 oz)