Phadia 100 - User Manual
Maintenance
12-3501-20/12
Page 11.21
2
5
6
11
Secondary Fuses
1.
Disconnect the mains
power cord.
2.
Use a screwdriver to open
the fuse holder. Put the screw-
driver in the vertical slot in the
fuse holder and turn it counter-
clockwise.
3.
Take out the fuse holder
from the instrument and re-
move the old fuse.
4.
Place a new fuse in the
fuse holder and install it in the
instrument.
5.
Connect the mains power
cord.
Replace broken Bottle Holders on the
Sample Carousel
1.
Push the Bottle holder
back towards the edge of the
Sample Carousel. Lift the front
of the Bottle holder and turn
the holder sideways. Pull the
Bottle holder up. Fig. 1.
2.
Place a new Bottle holder
in the position where the bro-
ken one was placed.
NOTE! Phadia 100 must only be connected to a
grounded power outlet (protective ground).
Fig.1