Phadia 100 - User Manual
Assay Run Editing Functions
12-3501-20/12
Page 4.3
3
4
5
6
4
Change Dilution for Sample
1.1.Enter Request
[24]
Enter Dilution Factor:1_
1.1.Enter Request
[24]
Sample Dilution:INSTRUMENT DILUTION?
1.1.Enter Request
[24]
Enter Sample ID:3627_
1.1.Enter Request
[24]
Sample Dilution:PREDILUTED?
The following steps describes how you change the replicate and dilution for a
sample when you are in Enter Request in the Prepare Run mode.
1.
You can change the dilu-
tion factor for a sample by typ-
ing the sample identity and then
pressing
shift
and
enter
keys
at the same time. In this case all
tests on this sample will be di-
luted.
2.
Press
yes
if you want to
add a prediluted sample, other-
wise press
no
. If yes go to step
4.
If predilution is selected the dilution factor can be 1-1000.
3.
Press
yes
if you want the
instrument to perform the dilu-
tion, otherwise
no
.
If you select instrument dilution the dilution factor can only be 1, 2, 5, 10, 20,
50 or 100.
4.
Type the dilution factor
and press
enter
.
5.
Enter the test names or
test panel names by typing them
one at a time and press
enter
to
confirm each test name.
6.
To save the entered tests
for the sample identity, press
enter
one more time.
Regarding dilutions, see also chapter 3, Reagent and Sample Handling and the
corresponding Directions for Use.
3627
(sIgE) Test:_ [24]
3627 (sIgE) Test:f1_ [21]
t3, w1, g6
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...