Error Handling
Phadia 100 - User Manual
P a g e 1 2 . 4 2
12-3501-20/12
Example from a Laboratory Report:
For further information regarding Curve Controls, see chapter 7, Monthly
Calibration, and regarding the Laboratory Report, see chapter 10, Printouts.
CC-1 HIGH and CC-2 HIGH
What can cause the problem:
No Washing Solution going through Wash Rake, or lack of Stop Solution. For
ImmunoCAP Specific IgA, ImmunoCAP Specific IgG
and
EliA
, contamination
problems (most likely contamination of conjugate or CC vial with sample). For
EliA
, wrong Lot Specific Code, no Washing Concentrate in Washing Solution or
using Specific IgG Conjugate can cause this error. Running
Total IgE method
using
Specific IgE Conjugate
, can give this error message. For
Specific IgE
method wrong Curve Controls can have been used.
Corrective actions:
Follow instructions in Action List in order to identify or eliminate the errors. For
additional information see Explanation of Actions, number
3, 7, 18, 22, 24, 25,
26, 27, 30, 32.
If the error remains contact your local Phadia representative.
CC-1 HIGH and CC-2 OK
What can cause the problem:
Insufficient wash or use of wrong Curve Control. For
ImmunoCAP Specific IgA,
ImmunoCAP Specific IgG, ImmunoCAP Specific IgG4
and
EliA
it can also be a
contamination problem (most likely contamination of conjugate or CC vial with
sample).
CURVE CONTROLS
NAME----RESP--%CV--CONC--%CV--JUDGED
CC-1
4696
0.51
HIGH
CC-2
988
0.096
LOW
CALIBRATION CURVE IS NOT OK
THE SAMPLE RESULTS ARE PROBABLY NOT OK
SEE ACTION LIST FOR CORRECTIVE ACTIONS
AND/OR JUDGEMENT OF RESULTS.
OK, HIGH*, LOW*, HIGH or LOW
is printed here
The Curve Control Error is printed
here
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...