Phadia 100 - User Manual
Error Handling
12-3501-20/12
Page 12.41
12
ImmunoCAP Specific IgG, ImmunoCAP Specific IgG4
and
EliA
, contamination
problems.
Corrective actions:
Follow instructions in Action List in order to identify or eliminate the errors. For
additional information see Explanation of Actions, number
7, 22, 24, 38.
If the error remains contact your local Phadia representative.
CALIBRATION CURVE OLDER THAN RECOMMENDED
What can cause the problem:
The Calibration Curve is too old.
Corrective actions:
Run a new Calibration Curve in the next assay run.
Curve Control Errors
Curve Control Errors occur when Curve Controls are outside the inner limits of
the Calibration Curve.
If the Curve Controls are outside the inner limits, you get HIGH*, HIGH, LOW*
or LOW under Curve Controls on the Laboratory Report.
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...