Phadia 100 - User Manual
Routine Assay Run
12-3501-20/12
Page 2.37
2
3
4
5
6
2
13
1.1.Enter Request
[36]
Enter Sample ID:_
1.1.Enter Request
[6]
CALIBRATORS ADDED TO RUN
Messages that can occur
CALIBRATION CURVE ALREADY EXISTS. RUN NEW?
There are already calibrators for the method you selected. If you press
yes
you
will add new calibrators to the run. If you press
no
, the existing ones will be used.
CALIBRATORS ALREADY ADDED
You have already added calibrators to this method and run.
TOO MANY METHODS AT SAME TIME
The number of methods possible to enter was exceeded. Maximal number of
methods that can be entered in one run is four (4).
NOT ENOUGH ROOM FOR CALIBRATORS
There is not enough ImmunoCAP/EliA Well or sample positions left to add the
calibrators to the run.
METHOD NOT COMPATIBLE
El-G method is not compatible with methods using ImmunoCAP.
5.
When you have selected
the method you want, there will
be a message (which will be
shown for just a few seconds)
telling you that the calibrators
are added to the run.
6.
Now you can continue
adding sample identities. If you
want to enter patient samples
go to Load and Start.
7.
To return to previous
mode press
mode
or
enter
.
Then you will return to the Pre-
pare Run mode.
For information how to print a Specific Run with selected calibrators, see
chapter 10, Printouts.
Phadia 100 Ver.3.0
13:48:33
1.Prepare Run?
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...