Printouts
12-3501-20/12
Page 10.13
10
QUALITY CONTROL
ImmunoCAP
Specific IgE 0-100
Identity:
Specific IgE Control
t1
Lot number:
AVMAB
Expected conc: 0.70
Expected s:
0.05
Limit s:
3.0
LOG GRAPH
Concentration (kUA/l)
0.85
0.70
0.55
090512
090514
090516
090518
LOG DATA
RUN DATE,
TIME---LOT NO---CAL CODE--CONCENTRATION
090512 11:14
AVMAB 1F
0.72
0.65
090513 11:18
AVMAB 1T
0.66
0.68
090514 12:58
AVMAB 1T
0.61
0.69
090515 13:02
AVMAB 2T
0.70
0.69
090516 11:05
AVMAB 2T
0.73
0.69
090517 12:33
AVMAC 2T
0.72
0.75
090518 12:53* AVMAC 2T
<0.35
0.78
090519 12:43
AVMAC 2T
0.77
0.74
LOT NO------MEAN-------s-----%CV---
AVMAB
0.68
0.03
5.1
AVMAC
0.75
0.02
2.8
Quality Control Log
This printout lists the data logged for one Quality Control point and a graph of
the logged data. To print the Quality Control log select
4.1.1.Quality control-
log
on the instrument. The latest 100 logs are stored.
Date and time when each point
was logged
Lot number of each logged point
as stated on QC vial
The asterisc indicates that a point
has been inactivated and is not
plotted in the graph or included
in the mean, standard deviation
or % CV
Calculated mean concentration
and standard deviation for all QC
points with the same lot no
Lot number for QC
Test name of the QC
0354, 090520 10:12, 2.3, 032
If a start date for the print is
defined, it is noted here
Instrument serial number
Method name
Data listed here depends on how
the limits are specified in the
method
Identity of the QC
Date and time of printout
Software version number
Method version number
Concentration of each logged
point, in this example two repli-
cates
Calibration code of used conju-
gate for each logged point
Expected concentration as de-
fined in the method
Expected standard deviation as
defined in the method
Maximum allowed standard de-
viation as defined in the method
Graphic presentation of the
logged data for the selected
lotnumber. Each replicate is in-
cluded. Inactivated points are not
included. The format of the graph
depends on limits defined in the
method.
•
•
•
••
•
•
•
•
•
•
•
•
•
•
•
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...