Phadia 100 - User Manual
Quality Assurance
12-3501-20/12
Page 8.13
8
Summary Reports
In addition valuable summaries, called Summary reports are sent out after each
4-month cycle covering the last three cycles (12 month).
A 12 monthly trend statistics reflecting the long-term method performance in
the laboratory is also distributed. These data helps the laboratories to maintain
test procedures or improve them where necessary. They provide timely signals
of system errors, calibration problems or can indicate a need to improve
technican training.
For each laboratory the report is based on at most 36 (12x3) observations for
Specific IgE and EliA Autoimmunity, and 24 (12x2) observations for Total IgE
and ECP.
Any questions regarding the interpretation of the summary reports, please
contact your local sales office or Phadia AB.
Results from all labs
Cat dander (e1)
N (except outliers)
525
No of participating labs: 530
Median
28,9 kU
A
/l
Mean
28,8 kU
A
/l
SD
2,52 kU
A
/l
CV
8,74 %
Mean - 2 SD
23,8 kU
A
/l
Mean + 2 SD
33,9 kU
A
/l
Your results
29,8 kU
A
/l
Deviation from mean
3,34 %
Deviation score
0,38
0
5
10
15
20
25
Conc. of Specific IgE (kU
A
/l)
P
e
rc
e
n
t o
f a
ll
lab
s
Time Trend Plot
-4
-3
-2
-1
0
1
2
3
4
200501
200502
200503
200504
200505
200506
200507
200508
200509
200510
200511
200512
Month
D
evi
at
io
n
S
c
o
re
Allergen 1
Allergen 2
Allergen 3
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...