Instrument Specification and Installation
Phadia 100 - User Manual
P a g e 1 . 2 8
12-3501-20/12
4.
Press
yes
.
5.
Enter the TCP/IP Address,
then press
Enter
.
6.
Press
Enter
to return to
the selection menu.
7.
Press
no
until 7 is dis-
played.
8.
Press
yes
.
9.
Enter the TCP/IP Port (de-
fault port is 5004), then press
Enter
.
Configure computer name
When using more than one Phadia 100 in the same network, the computer name
has to be set. To enter this menu you need access to the laboratory password.
Follow the steps below when you set the computer name.
1.
In Install/Service mode
press
8
.
2.
Press
yes
.
3.
Enter the computer name
(in this example BARU1431),
then press
Enter
.
6.7.6.IDM TCP/IP Address
Enter Address:
6.7.6.IDM TCP/IP Address
Enter Address:192.168.0.10
6.7.Communication parameters
COM2
7.IDM TCP/IP Port?
6.7.6.IDM TCP/IP Port
Enter Port:
6.7.6.IDM TCP/IP Port
Enter Port: 5004
6.7.Communication parameters
COM2
8.Computer name?
6.7.8.Computer
Computer name:
6.7.8.Computer
Computer name: BARU1431
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...