Routine Assay Run
Phadia 100 - User Manual
P a g e 2 . 4 8
12-3501-20/12
EXPIRY DATE PASSED, CONTINUE?
This question will appear on the display if bar-code is read on a reagent if the
expiry date is passed. Press
yes
to use anyway or
no
to use another reagent.
THIS REAGENT IS NOT REQUESTED
This message will appear if bar-code is read on a reagent not requested for this
run.
THIS SAMPLE IS NOT REQUESTED
This message will appear if bar-code is read on a sample not requested for this run.
THIS ImmunoCAP IS NOT REQUESTED
This message will appear if bar-code is read on an ImmunoCAP carrier not
requested for this run.
THIS EliA WELL IS NOT REQUESTED
This message will appear if bar-code is read on an EliA Well tube not requested
for this run.
REMOVE EMPTY CARRIER
The ImmunoCAP/EliA Well carrier is empty, but more ImmunoCAP/Elia Well
of this type is required.
SAMPLE CAROUSEL COVER IS NOT PRESENT
Put the Sample Carousel Cover in position.
PROCESSING LID IS NOT CLOSED
Close the Processing Lid.
PROCESSING LID IS NOT LOCKED
Make sure that the Processing Lid is closed properly and that nothing is blocking
the lid.
WRONG BAR-CODE IDENTITY
This message will appear if bar-code is read on something other than reagent,
sample or ImmunoCAP. Check that you have not read the EAN code instead of
the article bar-code.
WRONG CALIBRATION CODE
This message will appear if bar-code is read on a ImmunoCAP carrier with
wrong calibration code. Find the ImmunoCAP Carrier or run a new calibration
curve.
REMOVE CARRIER
This message will appear when to remove ImmunoCAP/EliA Well Carrier.
CHECK CAPS ON VIALS
Check that there are no caps left on vials.
ALL LOADED. REMOVE CARRIER
All ImmunoCAP are loaded. Remove the ImmunoCAP/EliA Well Carrier.
BLANK CHECK ERROR
For corrective actions see Action list.
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...