Phadia 100 - User Manual
Assay Run Editing Functions
12-3501-20/12
Page 4.21
3
4
5
6
4
10.
Type the Identity of the
Quality Control (up to eight
characters) and press
enter
to
confirm this step.
Messages that can occur
SAME QUALITY CTRL IDENTITY ALREADY EXISTS
The identity is already used by another QC.
11.
Type the test name and
press
enter
to confirm this step.
Messages that can occur
ENTERED TEST DOES NOT EXIST
The test name does not exist for the method. Please, check the spelling.
SAME QC IDENTITY ALREADY EXISTS
The identity is already used by another QC.
12.
Press
1
,
2
,
3
,
4
or
5
de-
pending on how many repli-
cates you want to run. (1 is
default.) Press
enter
to con-
firm this step.
Messages that can occur
Replicate: TOO HIGH!
The number of replicates is too high. Please, choose a number between 1 and 5.
Replicate: TOO LOW!
The number of replicates is too low. Please, choose a number between 1 and 5.
5.2.9.1.Add QC
[sIgE]
Test name:
5.2.9.1.Add QC
[sIgE]
Replicate:1
5.2.9.1.Add QC
[sIgE]
Identity:
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...