5-14 Image Optimization
Effects
This function is used to adjust the scan angle of linear probes, so as to change
the angle between the transmitting beam and flow direction.
Impacts
Steer is only valid for linear probes.
Line Density
Description
Line density determines the quality and information of the image.
Operations
Adjust through [Line Density] on the menu.
There are four levels of line density available: UH, M, H, L.
Effects
The higher the line density, the higher the resolution.
Impacts
The higher the line density, the lower the frame rate.
Packet Size
Description
This function is an indication of the ability to detect flow, which is used to adjust
the accuracy of color flow.
Operations
Select different effects through [Packet Size].
There are 3 levels of packet size available: 0 represents no packet size control
and the bigger the value the higher the sensitivity.
Effects
The higher the packet size, the more sensitive the indication for low-velocity flow.
Impacts
Adjusting the packet size may lead to the frame rate changing.
Flow State
Description
This function is used for fast image optimization.
Operations
Select different effects through [Flow State].
3 levels are provided: L, M, H.
Persistence
Description
This function adjusts the temporal smoothing in Color mode to optimize the
image.
Operations
Select different effects through [Persistence].
The system provides 6 levels of persistence adjustment: 0 represents no
persistence, and the bigger the value the stronger the effect.
Smooth
Description
This feature is used to reject noise and smooth the image.
Operations
Adjust through [Smooth] on the menu.
The system provides 6 levels of smooth function: the bigger the value the
stronger the effect.
Summary of Contents for TE5
Page 1: ...TE7 TE5 Diagnostic Ultrasound System Operator s Manual Basic Volume ...
Page 2: ......
Page 6: ......
Page 12: ......
Page 24: ......
Page 36: ......
Page 54: ......
Page 110: ......
Page 115: ...Display Cine Review 6 5 6 Tap Return on the screen or tap Freeze to exit image compare ...
Page 120: ......
Page 124: ......
Page 156: ......
Page 174: ......
Page 192: ...12 18 Setup Select Advanced and do as follows Select MAPS and do as follows ...
Page 202: ...13 2 Probes and Biopsy C5 2s L12 4s L7 3s P4 2s L14 6s C11 3s L14 6Ns V11 3Ws P7 3Ts 7LT4s ...
Page 226: ...13 26 Probes and Biopsy NGB 034 NGB 035 ...
Page 250: ......
Page 272: ......
Page 276: ...A 4 Wireless LAN Tap Add Manually create a network profile to set ...
Page 282: ......
Page 318: ......
Page 322: ......
Page 323: ...P N 046 006959 07 1 0 ...