Safety Precautions 1-9
9.
The probe sheath contains natural rubber that can cause allergic
reactions in some individuals.
10.
Do not use pre-lubricated condoms as a sheath. Lubricant may
not be compatible with the transducer material and damage may
result.
11.
Transducer damage may be caused by inappropriate gel,
detergent or cleanser:
Do not soak or saturate transducers with solutions containing
alcohol, bleach, ammonium chloride compounds, acetone or
formaldehyde.
Avoid contact with solutions or coupling gels containing mineral
oil or lanolin.
12.
The contrast agent used must comply with the relevant local
regulations.
NOTE:
1. Read the following precautions to prevent the probe from malfunction:
Clean and disinfect the probe before and after each examination.
After the examination, wipe off the ultrasound gel thoroughly. Otherwise,
the ultrasound gel may solidify and the image quality would be degraded.
Summary of Contents for TE5
Page 1: ...TE7 TE5 Diagnostic Ultrasound System Operator s Manual Basic Volume ...
Page 2: ......
Page 6: ......
Page 12: ......
Page 24: ......
Page 36: ......
Page 54: ......
Page 110: ......
Page 115: ...Display Cine Review 6 5 6 Tap Return on the screen or tap Freeze to exit image compare ...
Page 120: ......
Page 124: ......
Page 156: ......
Page 174: ......
Page 192: ...12 18 Setup Select Advanced and do as follows Select MAPS and do as follows ...
Page 202: ...13 2 Probes and Biopsy C5 2s L12 4s L7 3s P4 2s L14 6s C11 3s L14 6Ns V11 3Ws P7 3Ts 7LT4s ...
Page 226: ...13 26 Probes and Biopsy NGB 034 NGB 035 ...
Page 250: ......
Page 272: ......
Page 276: ...A 4 Wireless LAN Tap Add Manually create a network profile to set ...
Page 282: ......
Page 318: ......
Page 322: ......
Page 323: ...P N 046 006959 07 1 0 ...