System Maintenance 17-5
Compatible Disinfectants
The following table lists compatible disinfectants.
Manufacturer
Product Name
Type
Advanced Ultrasound Solutions Inc.
SONO
TM
ULTRASOUND WIPES
Wipe
Parker laboratories Inc
Protex
TM
Disinfectant Spray
Spray
PDI Inc.
Sani-Cloth® Plus germicidal
disposable cloth
Wipe
Bode Chemie Gmbh
Mikrobac Tissues
Wipe
GAMA Healthcare Ltd.
Clinell universal wipes
Wipe
PDI Inc.
Sani-Cloth® AF3
Wipe
Diversey
VIREX II 256
Solution
Diversey
VIREX TB
Solution
Schulke
Schulke mikrozid
®
Sensitive
Wipes
Wipe
/
0.5% sodium hypochlorite
Solution
/
75% alcohol
Solution
Disinfection Procedures
NOTE:
Clean the system in power off status; or in power on status, tap
in the top-right
corner of the screen and select
to lock the system for 10 seconds.
The ultrasound system has passed puncture test and leakage current test, and thus
can be disinfected on power-on status.
After cleaning, disinfect the main unit as follows.
1.
Wear medical gloves to prevent infection.
2.
Disinfect the main unit with disinfectant wipe, spray or solution. Follow the disinfectant
manufacturer’s recommended contact time and method.
NOTE: Do not disinfect handle, or any visible sockets or interfaces (such as probe sockets,
ventilation holes, dust-proof cover, loudspeaker, sockets or interfaces in the IO panel and
power supply panel). See the following figure.
Summary of Contents for TE5
Page 1: ...TE7 TE5 Diagnostic Ultrasound System Operator s Manual Basic Volume ...
Page 2: ......
Page 6: ......
Page 12: ......
Page 24: ......
Page 36: ......
Page 54: ......
Page 110: ......
Page 115: ...Display Cine Review 6 5 6 Tap Return on the screen or tap Freeze to exit image compare ...
Page 120: ......
Page 124: ......
Page 156: ......
Page 174: ......
Page 192: ...12 18 Setup Select Advanced and do as follows Select MAPS and do as follows ...
Page 202: ...13 2 Probes and Biopsy C5 2s L12 4s L7 3s P4 2s L14 6s C11 3s L14 6Ns V11 3Ws P7 3Ts 7LT4s ...
Page 226: ...13 26 Probes and Biopsy NGB 034 NGB 035 ...
Page 250: ......
Page 272: ......
Page 276: ...A 4 Wireless LAN Tap Add Manually create a network profile to set ...
Page 282: ......
Page 318: ......
Page 322: ......
Page 323: ...P N 046 006959 07 1 0 ...