Probes and Biopsy 13-15
Do not soak the probe connector or the cable near it into water or any solution.
Follow local regulations when selecting and using the sterilization solution.
3. Rinse the probe with plenty of sterile water (about 2 gallons) for at least 1 minute to remove all
chemical residues on it. Or, follow the rinsing method recommended by the sterilization
solution manufacturer to rinse the probe.
4. Wipe off the water on the probe with sterile cloth or gauze after rinsing it. Do not dry the probe
by heating.
Immerse the probe in the solution (take 7LT4s as an illustration)
NOTE:
Repeated disinfection may degrade the performance and safety of the probe.
13.1.6 Storage and Transportation
When all examinations for the session have been completed, confirm that the probe is in good
condition. After disinfecting the probe, confirm that the probe is still in good condition and stored in
a suitable place.
1. To prevent the probe from being damaged, DO NOT store it where it may be exposed to:
Direct sunlight or X-rays
Sudden changes in temperature
Dust
Excessive vibration
Heat generators
2. Store and transport the probe (L12-4s, L14-6Ns, V11-3Ws, SC6-1s, SP5-1s, 7L4s, L9-3s, L11-
3VNs, L12-3RCs, L14-5Ws and C11-3s) under the following ambient conditions:
Ambient temperature: -20°C to 55°C
Relative humidity: 20% to 95% (no condensation)
Atmospheric pressure: 700 hPa to 1060 hPa
Store and transport the probe (C5-2s, L7-3s, L14-6s, P4-2s, P10-4s, P7-3s, 6CV1s, C5-1s and
7LT4s) under the following ambient conditions:
Ambient temperature: -20°C to 55°C
Relative humidity: 30% to 95% (no condensation)
Atmospheric pressure: 700 hPa to 1060 hPa
Store and transport the P7-3Ts probe under the following ambient conditions:
Ambient temperature: -10°C to 45°C
Relative humidity: 30% to 90% (no condensation)
Atmospheric pressure: 700 hPa to 1060 hPa
Store and transport the L14-5sp probe under the following ambient conditions:
Ambient temperature: 0°C to 60°C
Relative humidity: 30% to 95% (no condensation)
Atmospheric pressure: 700 hPa to 1060 hPa
Store and transport the L20-5s probe under the following ambient conditions:
Summary of Contents for TE5
Page 1: ...TE7 TE5 Diagnostic Ultrasound System Operator s Manual Basic Volume ...
Page 2: ......
Page 6: ......
Page 12: ......
Page 24: ......
Page 36: ......
Page 54: ......
Page 110: ......
Page 115: ...Display Cine Review 6 5 6 Tap Return on the screen or tap Freeze to exit image compare ...
Page 120: ......
Page 124: ......
Page 156: ......
Page 174: ......
Page 192: ...12 18 Setup Select Advanced and do as follows Select MAPS and do as follows ...
Page 202: ...13 2 Probes and Biopsy C5 2s L12 4s L7 3s P4 2s L14 6s C11 3s L14 6Ns V11 3Ws P7 3Ts 7LT4s ...
Page 226: ...13 26 Probes and Biopsy NGB 034 NGB 035 ...
Page 250: ......
Page 272: ......
Page 276: ...A 4 Wireless LAN Tap Add Manually create a network profile to set ...
Page 282: ......
Page 318: ......
Page 322: ......
Page 323: ...P N 046 006959 07 1 0 ...