5-4 Image Optimization
5.2
B Mode Image Optimization
B mode is the basic imaging mode that displays real-time images of anatomical tissues and
organs.
5.2.1 Basic Procedures for B Mode Imaging
1. Enter the patient information. Select an appropriate probe and exam mode.
2. Tap [B] on the right side of the operating panel to enter B mode.
3. Tap [Image] to open the image menu. Adjust the parameters to optimize the image.
4. Perform other operations (e.g., measurement and calculation) if necessary.
Tip: tap [B] to return to B mode at any time (except LVO mode).
5.2.2 B Mode Parameters
In B mode scanning, the image parameter area in the top-left corner of the screen displays the
real-time parameter values as follows:
Parameter
F
D
G
FR
DR
Meaning
Frequency Depth Gain Frame Rate Dynamic Range
5.2.3 B Mode Image Optimization
Frequency (Image Quality)
Description
To switch between the fundamental frequency and harmonic frequency as well
as select the corresponding frequency type. The real-time frequency value is
displayed in the image parameter area in the top-left corner of the screen, and if
harmonic frequency is used “F H” is displayed as the harmonic frequency value.
Select the different frequency values through
at the left part of the image
area.
The adjusting range of the harmonic frequency values can be divided into 4
levels: penetration preferred (HPen), general mode (HGen), resolution preferred
(HRes), and the mode between penetration preferred and general mode (HPen-
Gen).
The adjusting range of fundamental frequency values can be divided into 3
levels: penetration preferred (Pen), general mode (Gen), and resolution
preferred (Res).
Impacts
The system provides an imaging mode using harmonics of echoes to optimize
the image. Harmonic imaging enhances near-field resolution and reduces low-
frequency and large amplitude noise, so as to improve small parts imaging.
Select the frequency according to the detection depth and current tissue
features.
Gain
Description
To adjust the gain of the whole receiving information in B mode. The real-time
gain value is displayed in the image parameter area in the top-left corner of the
screen.
Summary of Contents for TE5
Page 1: ...TE7 TE5 Diagnostic Ultrasound System Operator s Manual Basic Volume ...
Page 2: ......
Page 6: ......
Page 12: ......
Page 24: ......
Page 36: ......
Page 54: ......
Page 110: ......
Page 115: ...Display Cine Review 6 5 6 Tap Return on the screen or tap Freeze to exit image compare ...
Page 120: ......
Page 124: ......
Page 156: ......
Page 174: ......
Page 192: ...12 18 Setup Select Advanced and do as follows Select MAPS and do as follows ...
Page 202: ...13 2 Probes and Biopsy C5 2s L12 4s L7 3s P4 2s L14 6s C11 3s L14 6Ns V11 3Ws P7 3Ts 7LT4s ...
Page 226: ...13 26 Probes and Biopsy NGB 034 NGB 035 ...
Page 250: ......
Page 272: ......
Page 276: ...A 4 Wireless LAN Tap Add Manually create a network profile to set ...
Page 282: ......
Page 318: ......
Page 322: ......
Page 323: ...P N 046 006959 07 1 0 ...