Image Optimization 5-23
Trace
Smooth
Adjust through [Trace Smooth] on the menu.
There are 4 levels of smooth effect provided, the bigger the value, the higher the
smooth processing.
Trace
Sensitivity
This function is used to set the sensitivity of tracing in the spectrum.
Adjust through [Trace Sensitivity] on the menu.
There are 5 levels of sensitivity adjustment, the bigger the value the higher the
sensitivity.
Operations
In real-time scanning, the results displayed are derived from the calculation of
the latest cardiac cycle.
In the freeze and cine status, the results displayed are calculated from the
current selected area.
Invert
Description
This function is used to set how the spectrum is displayed.
Operations
Turn the function on or off using the [Invert] control.
Select “Auto Invert” in the [Setup] -> [System] -> [Image], so the spectrum can
automatically invert when the color flow is steered to a certain angle to
accommodate the operator’s desire to distinguish the flow direction.
Speed
Description
This function is used to set the scanning speed of PW mode imaging.
Operations
Adjust through [Speed (mm/s)] on the menu.
There are 6 levels of scan speed available.
Effects
Changing the speed makes it easier to identify the cardiac cycles and to detect
more details.
T/F Res.
Description
This function is used to create a balance between time resolution and spatial
resolution.
Operations
Adjust through [T/F Res.] on the menu.
There are 4 levels of T/F Res. values available.
Wall Filter
Description
It filters out low-velocity signals to provide effective information, and this function
is used to adjust the filtered frequency. The real-time value is displayed in the
image parameter area in the top-left corner of the screen.
Operations
Adjust through [WF] on the menu.
7 levels of wall filter function are provided.
Impacts
Signals of low-velocity flow may be missing.
Summary of Contents for TE5
Page 1: ...TE7 TE5 Diagnostic Ultrasound System Operator s Manual Basic Volume ...
Page 2: ......
Page 6: ......
Page 12: ......
Page 24: ......
Page 36: ......
Page 54: ......
Page 110: ......
Page 115: ...Display Cine Review 6 5 6 Tap Return on the screen or tap Freeze to exit image compare ...
Page 120: ......
Page 124: ......
Page 156: ......
Page 174: ......
Page 192: ...12 18 Setup Select Advanced and do as follows Select MAPS and do as follows ...
Page 202: ...13 2 Probes and Biopsy C5 2s L12 4s L7 3s P4 2s L14 6s C11 3s L14 6Ns V11 3Ws P7 3Ts 7LT4s ...
Page 226: ...13 26 Probes and Biopsy NGB 034 NGB 035 ...
Page 250: ......
Page 272: ......
Page 276: ...A 4 Wireless LAN Tap Add Manually create a network profile to set ...
Page 282: ......
Page 318: ......
Page 322: ......
Page 323: ...P N 046 006959 07 1 0 ...