Image Optimization 5-25
Angle
Description
This function is used to adjust the angle between Doppler vector and flow to
make the velocity more accurate.
The real-time adjusting angle value is displayed on the left part of the spectrum
map.
Operations
Adjust through [Angle] on the menu.
Tap [-] or [+] to change the value slightly or drag the control directly.
The adjustable angle range is -89~89°, in increments of 1°.
Quick Angle
Description
To adjust the angle faster, in increments of 60°. The real-time value is displayed
on the left part of the spectrum map.
Operations
Tap the three buttons above [Quick Angle] on the bottom of the image area.
There are 3 angles for quick adjustment: -60°, 0° and 60°.
Dynamic Range
Description
The dynamic range conveys the information which is being transformed from
echo intensity to grayscale.
Operations
Adjust dynamic range through [Dyn Ra.].
Tap [-] or [+] to change the value slightly or drag the control directly.
The adjusting range is 24-72.
Effects
The more the dynamic range, the more specific the information and the lower the
contrast with more noise.
PW Volume
Description
This function is used to adjust the output audio in the spectrum map.
Operations
Adjust through [Volume] on the menu.
Tap [-] or [+] to change the value slightly or drag the control directly.
The adjusting range of the audio is 0-100%.
Effects
Utilizing the output audio helps to identify the feature and the status of flow.
PW Steer
Description
This function is used to adjust the angles for the sampling line.
Operations
Steer the probe by tapping buttons on the bottom of the image area.
Effects
This feature is used to steer the direction of the beam so as to change the angle
between the beam and flow direction with immobility of the linear probe.
Impacts
The PW Steer function is available only for linear probes.
Summary of Contents for TE5
Page 1: ...TE7 TE5 Diagnostic Ultrasound System Operator s Manual Basic Volume ...
Page 2: ......
Page 6: ......
Page 12: ......
Page 24: ......
Page 36: ......
Page 54: ......
Page 110: ......
Page 115: ...Display Cine Review 6 5 6 Tap Return on the screen or tap Freeze to exit image compare ...
Page 120: ......
Page 124: ......
Page 156: ......
Page 174: ......
Page 192: ...12 18 Setup Select Advanced and do as follows Select MAPS and do as follows ...
Page 202: ...13 2 Probes and Biopsy C5 2s L12 4s L7 3s P4 2s L14 6s C11 3s L14 6Ns V11 3Ws P7 3Ts 7LT4s ...
Page 226: ...13 26 Probes and Biopsy NGB 034 NGB 035 ...
Page 250: ......
Page 272: ......
Page 276: ...A 4 Wireless LAN Tap Add Manually create a network profile to set ...
Page 282: ......
Page 318: ......
Page 322: ......
Page 323: ...P N 046 006959 07 1 0 ...