5-42 Image Optimization
Description of parameters:
Parameter
Description
Method
Function: select the image acquisition method.
Selection: Rocked, Linear.
Linear mode: during the sweep, the probe must be kept parallel. The
scanning speed should be constant.
Rocked mode: in this mode, the probe must be moved to a position
where you can clearly see a middle cut of the object you want to scan
and render. Tilt the probe to about 30 degrees until the object you
want to scan disappears. Start the acquisition and tilt the probe over a
distance of around 60 degrees until the object disappears again.
During the sweep, the probe may not be moved parallel, just tilted.
Tip: the speed is related to scanning distance or angle.
Distance
Function: to set the distance the probe covered from one end to the
other end during a linear sweep.
Range: 10-200 mm, in increments of 10 mm.
Angle
Function: to set the angle the probe covered during a fan sweep.
Range: 10-80°, in increments of 2°.
Render
Mode
Surface
Function: set Surface as the 3D image rendering mode.
This is useful for surface imaging, such as fetus face, hand or foot.
Tip: you may have to adjust the threshold to obtain a clear body
boundary.
Max.
Function: set Max. as the 3D image rendering mode. Displays the
maximum echo intensity in the observation direction.
This is useful for viewing bony structures.
Min.
Function: set Min. as the 3D image rendering mode. Displays the
minimum echo intensity in the observation direction.
This is useful for viewing vessels and hollow structures.
X-ray
Function: set X-ray as the 3D image rendering mode. Displays the
average value of all gray values in the ROI.
X Ray: used for imaging tissues with different internal structures or
tissues with tumors.
5.11.2.3 Smart 3D Image Viewing
Enter/Exit Image Viewing
To enter image viewing:
The system enters image viewing when image acquisition is complete.
Exit
To return to Smart 3D image acquisition preparation status, tap [Update] or [Freeze].
Activate MPR
Tap [A], [B], [C] or [VR] to activate MPR or 3D image (VR).
Summary of Contents for TE5
Page 1: ...TE7 TE5 Diagnostic Ultrasound System Operator s Manual Basic Volume ...
Page 2: ......
Page 6: ......
Page 12: ......
Page 24: ......
Page 36: ......
Page 54: ......
Page 110: ......
Page 115: ...Display Cine Review 6 5 6 Tap Return on the screen or tap Freeze to exit image compare ...
Page 120: ......
Page 124: ......
Page 156: ......
Page 174: ......
Page 192: ...12 18 Setup Select Advanced and do as follows Select MAPS and do as follows ...
Page 202: ...13 2 Probes and Biopsy C5 2s L12 4s L7 3s P4 2s L14 6s C11 3s L14 6Ns V11 3Ws P7 3Ts 7LT4s ...
Page 226: ...13 26 Probes and Biopsy NGB 034 NGB 035 ...
Page 250: ......
Page 272: ......
Page 276: ...A 4 Wireless LAN Tap Add Manually create a network profile to set ...
Page 282: ......
Page 318: ......
Page 322: ......
Page 323: ...P N 046 006959 07 1 0 ...