E-4 Electrical Safety Inspection
LIMITS
For BF
applied parts:
100 μA in Normal Condition.
500 μA in Single Fault Condition.
E.8 Mains on Applied Part Leakage
The Mains on Applied Part test applies a test voltage, which is 110% of the mains voltage
using a limiting resistance, to selected applied part terminals. Current measurements are then
taken between the selected applied part and earth. Measurements are taken with the test
voltage (110% of mains) on applied parts in the normal and reverse polarity conditions.
The following outlet conditions apply when performing the Mains on Applied Part test.
Normal Polarity;
Reversed Polarity.
LIMITS
For BF
applied parts: 5000 μA.
E.9 Patient Auxiliary Current
Patient Auxiliary currents are measured between any selected Applied Part connector and the
remaining Applied Part connectors. All measurements may have a true RMS response.
The following outlet conditions apply when performing the Patient Auxiliary Current test.
normal polarity (Normal Condition);
reverse polarity (Normal Condition);
normal polarity with open neutral (Single Fault Condition);
reverse polarity with open neutral (Single Fault Condition);
normal polarity with open earth (Single Fault Condition);
reverse polarity with open earth (Single Fault Condition).
LIMITS
For BF
applied parts,
100 μA in Normal Condition.
500 μA in Single Fault Condition.
NOTE:
Make sure the safety analyzer is authorized and complies with the requirements of
IEC60601-1.
Follow the analyzer manufacturer's instructions.
Summary of Contents for TE5
Page 1: ...TE7 TE5 Diagnostic Ultrasound System Operator s Manual Basic Volume ...
Page 2: ......
Page 6: ......
Page 12: ......
Page 24: ......
Page 36: ......
Page 54: ......
Page 110: ......
Page 115: ...Display Cine Review 6 5 6 Tap Return on the screen or tap Freeze to exit image compare ...
Page 120: ......
Page 124: ......
Page 156: ......
Page 174: ......
Page 192: ...12 18 Setup Select Advanced and do as follows Select MAPS and do as follows ...
Page 202: ...13 2 Probes and Biopsy C5 2s L12 4s L7 3s P4 2s L14 6s C11 3s L14 6Ns V11 3Ws P7 3Ts 7LT4s ...
Page 226: ...13 26 Probes and Biopsy NGB 034 NGB 035 ...
Page 250: ......
Page 272: ......
Page 276: ...A 4 Wireless LAN Tap Add Manually create a network profile to set ...
Page 282: ......
Page 318: ......
Page 322: ......
Page 323: ...P N 046 006959 07 1 0 ...