Image Optimization 5-9
Echo Boost
Description
This function can improve contrast and decrease noise, so that a much clearer
boundary can be seen.
Operation
Set [Echo Boost] to be “On” to turn the function on.
Impacts
This function is only available using the phased probe in cardiac exam mode
(probe P7-3Ts does not support this function)
Patient Temperature
Description
If the current active probe is P7-3Ts, the parameter will display under the B mode
menu. You can enter the patient temperature by this function.
Operation
Enter the temperature by tapping [Patient Temperature].
WARNING:
If the patient temperature is above 37
°
C (98.6
°
F) and the [Patient
Temperature] setting is below the actual reading, the system could
overestimate the temperature of TEE transducer’s distal tip. This
could trigger the Auto-Cool feature. If the patient temperature
reaches or is near 37
°
C (98.6
°
F) and the [Patient Temperature]
setting is above the actual reading, the system could underestimate
the temperature of the TEE transducer's distal tip. The patient is
exposed to excessive temperatures.
Middle Line
Description
“Middle Line” helps to locate the focus point of shock wave during the treatment.
By watching the treatment procedure in real-time and adjusting the intension and
frequency of the shock wave, the harm to the patients can be reduced to the
least.
Operations
Tap [Middle Line] on the image menu to hide or display the line.
Dual Live
Description
Display different image effects of one probe for a better observation.
Operation
Tap [Dual Live] to turn on/off the function, and dual-split window of images are
displayed on the main screen.
Two pages of adjustable parameters are displayed; where, shared parameters
and left window parameters are displayed in the B(L) page, while right window
parameters are displayed in the B(R) page.
In the image parameter area on the top-left corner of the main screen,
parameters of the both windows are displayed.
Impacts
Image magnification is available in dual live mode.
Summary of Contents for TE5
Page 1: ...TE7 TE5 Diagnostic Ultrasound System Operator s Manual Basic Volume ...
Page 2: ......
Page 6: ......
Page 12: ......
Page 24: ......
Page 36: ......
Page 54: ......
Page 110: ......
Page 115: ...Display Cine Review 6 5 6 Tap Return on the screen or tap Freeze to exit image compare ...
Page 120: ......
Page 124: ......
Page 156: ......
Page 174: ......
Page 192: ...12 18 Setup Select Advanced and do as follows Select MAPS and do as follows ...
Page 202: ...13 2 Probes and Biopsy C5 2s L12 4s L7 3s P4 2s L14 6s C11 3s L14 6Ns V11 3Ws P7 3Ts 7LT4s ...
Page 226: ...13 26 Probes and Biopsy NGB 034 NGB 035 ...
Page 250: ......
Page 272: ......
Page 276: ...A 4 Wireless LAN Tap Add Manually create a network profile to set ...
Page 282: ......
Page 318: ......
Page 322: ......
Page 323: ...P N 046 006959 07 1 0 ...