6-2 Display & Cine Review
6.4
Freeze/Unfreeze the Image.
Tap [Freeze] in the bottom-right corner of the operating panel to freeze a scanning image. In freeze
mode, the probe stops transmitting acoustic power, and all images and parameters are frozen.
Tip: After freezing an image, the system may enter cine review, measure, comment adding, body
mark mode or current mode, depending on what has been preset. (Setting path: [Setup] ->
[System] -> [Image] -> “Freeze Config”)
Tap [Unfreeze] in freeze mode to unfreeze the image, and the system continues image scanning.
6.4.1 Imaging Mode Switching When Frozen
Imaging mode switching in freeze mode follows these principles:
In splitting display B mode, tap each image window to switch between the windows.
In freeze mode, the system supports imaging mode switching between the sub-modes
(only for the activated window). For example, if the frozen image is in B+Color+PW mode,
the system supports imaging mode switching between B+Color+PW, B+Color, B+PW and
B by tapping [Color] or [PW].
6.5
Cine Review
After tapping [Freeze], the system allows you to review and edit the images prior to the image
being frozen. This function is called cine review. You can perform zoom, measurements, add
comments and body marks on the images being reviewed.
The system supports manual review as well as automatic review.
In addition, the system supports the images reviewed along with physiological waveforms, if the
detection of physiological waveforms is performed.
CAUTION:
Cine Review images can be inadvertently combined in-between separate
patient scans. The Cine Review memory must be cleared at the end of the
current patient exam and before starting the next new patient exam by
tapping [End] in the bottom-left corner of the operating panel.
Cine files stored in the system’s hard drive should contain patient
information, to avoid the selection of an incorrect image file and potential
misdiagnosis.
6.5.1 Entering/Exiting Cine Review
To enter Cine Review:
Enter [Setup] -> [System] -> [Image] -> “Freeze Config” screen to set “Status after Freeze”
to “Cine.” The system enters manual cine review status once [Freeze] is tapped.
Open cine files in Review screen. The system enters automatic cine review status.
To exit Cine Review:
Tap [Freeze] or [B] and the system will return to image scanning and exit cine review.
Summary of Contents for TE5
Page 1: ...TE7 TE5 Diagnostic Ultrasound System Operator s Manual Basic Volume ...
Page 2: ......
Page 6: ......
Page 12: ......
Page 24: ......
Page 36: ......
Page 54: ......
Page 110: ......
Page 115: ...Display Cine Review 6 5 6 Tap Return on the screen or tap Freeze to exit image compare ...
Page 120: ......
Page 124: ......
Page 156: ......
Page 174: ......
Page 192: ...12 18 Setup Select Advanced and do as follows Select MAPS and do as follows ...
Page 202: ...13 2 Probes and Biopsy C5 2s L12 4s L7 3s P4 2s L14 6s C11 3s L14 6Ns V11 3Ws P7 3Ts 7LT4s ...
Page 226: ...13 26 Probes and Biopsy NGB 034 NGB 035 ...
Page 250: ......
Page 272: ......
Page 276: ...A 4 Wireless LAN Tap Add Manually create a network profile to set ...
Page 282: ......
Page 318: ......
Page 322: ......
Page 323: ...P N 046 006959 07 1 0 ...