Measurement 7-3
Gynecology
measurements - used for the uterus, ovary and follicles, etc.
Small Part measurements - used for small parts such as the thyroid, breast, MSK, superficial.
Urology measurements - used for the prostate, renal, micturated and testicular volume.
Vascular measurements - used for carotid, cerebral, upper and lower extremities vessels, etc.
EM measurements - used for EM-related full functional measurements.
Nerve measurements - used for nerve measurements.
7.4
Measurement Accuracy
Table 1 Basic Dimension Measurements
Parameter
Range
Error
Distance
Full Screen
Within ±4%
Area (ellipse, circle) Full Screen
Within ±7%
Trace Area
Full Screen
Within ±7%
Circ
Full Screen
Within ±10%
Angle
Full Screen
Within ±3%
Volume
Full Screen
Within ±12 %
Table 2 Basic Time Motion Measurements
Parameter
Range
Error
Distance
Full Screen
Within ±4 %
Time
Timeline Display
Within ±2%
Heart Rate
Timeline Display
Within ±5%
Velocity (PW mode)
10-200cm/s (for C5-2s, C11-3s,
L12-4s, L7-3s, L14-6s, L14-
6Ns, V11-3Ws, 7LT4s, P7-3Ts,
L14-5sp, L20-5s, 7L4s, SC6-1s,
6CV1s, L9-3s, C5-1s, C4-1s,
L11-3VNs, L12-3RCs, L14-
5Ws)
10-300cm/s (for P4-2s, P10-4s,
P7-3s, SP5-1s)
When angle≤60º, ≤5%.
Velocity (CW mode)
10-200cm/s (for P7-3Ts, C4-1s)
10-300cm/s (for P4-2s, P10-4s,
P7-3s, SP5-1s)
When angle≤60º, ≤5%. (not including
pencil probe)
NOTE:
Within the selected field range, measurement accuracy is ensured within the range
mentioned above. The accuracy specifications are performed in the worst conditions,
or based on real system tests, regardless of acoustic speed error.
Summary of Contents for TE5
Page 1: ...TE7 TE5 Diagnostic Ultrasound System Operator s Manual Basic Volume ...
Page 2: ......
Page 6: ......
Page 12: ......
Page 24: ......
Page 36: ......
Page 54: ......
Page 110: ......
Page 115: ...Display Cine Review 6 5 6 Tap Return on the screen or tap Freeze to exit image compare ...
Page 120: ......
Page 124: ......
Page 156: ......
Page 174: ......
Page 192: ...12 18 Setup Select Advanced and do as follows Select MAPS and do as follows ...
Page 202: ...13 2 Probes and Biopsy C5 2s L12 4s L7 3s P4 2s L14 6s C11 3s L14 6Ns V11 3Ws P7 3Ts 7LT4s ...
Page 226: ...13 26 Probes and Biopsy NGB 034 NGB 035 ...
Page 250: ......
Page 272: ......
Page 276: ...A 4 Wireless LAN Tap Add Manually create a network profile to set ...
Page 282: ......
Page 318: ......
Page 322: ......
Page 323: ...P N 046 006959 07 1 0 ...