Guidance and Manufacturer's Declaration 16-3
TABLE 2
GUIDANCE AND MINDRAY DECLARATION—ELECTROMAGNETIC IMMUNITY
The system is intended for use in the electromagnetic environment specified below. The
customer or the user of system should assure that it is used in such an environment.
IMMUNITY TEST
IEC 60601
TEST LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC
ENVIRONMENT-GUIDANCE
Electrostatic
Discharge(ESD)
IEC 61000-4-2
±6 kV contact;
±8 kV air
±6 kV contact;
±8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Electrical fast
Transient / burst
IEC 61000-4-4
±2 kV for power
supply lines;
±1 kV for
input/output lines
±2 kV for power
supply lines;
±1 kV for
input/output lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV line(s) to
line(s);
±2 kV line(s) to
earth
±1 kV line(s) to
line(s);
±2 kV line(s) to
earth
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, Short
interruptions and
voltage variation on
power supply input
voltage
IEC 61000-4-11
<5% U
T
(>95%
dip in U
T
) for 0.5
cycle
40% U
T
(60% dip
in U
T
) for 5 cycle
70% U
T
(30% dip
in U
T
) for 25
cycle
<5% U
T
(>95%
dip in U
T
) for 5
sec
<5% U
T
(>95% dip
in U
T
) for 0.5 cycle
40% U
T
(60% dip in
U
T
) for 5 cycle
70% U
T
(30% dip in
U
T
) for 25 cycle
<5% U
T
(>95% dip
in U
T
) for 5 sec
Mains power quality should be
that of a typical commercial or
hospital environment. If you
require continued operation
during power mains
interruptions, it is
recommended that our product
be powered from an
uninterruptible power supply or
a battery.
Power frequency
(50/60 HZ)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
NOTE:
U
T
is the A.C. mains voltage prior to application of the test level.
Summary of Contents for TE5
Page 1: ...TE7 TE5 Diagnostic Ultrasound System Operator s Manual Basic Volume ...
Page 2: ......
Page 6: ......
Page 12: ......
Page 24: ......
Page 36: ......
Page 54: ......
Page 110: ......
Page 115: ...Display Cine Review 6 5 6 Tap Return on the screen or tap Freeze to exit image compare ...
Page 120: ......
Page 124: ......
Page 156: ......
Page 174: ......
Page 192: ...12 18 Setup Select Advanced and do as follows Select MAPS and do as follows ...
Page 202: ...13 2 Probes and Biopsy C5 2s L12 4s L7 3s P4 2s L14 6s C11 3s L14 6Ns V11 3Ws P7 3Ts 7LT4s ...
Page 226: ...13 26 Probes and Biopsy NGB 034 NGB 035 ...
Page 250: ......
Page 272: ......
Page 276: ...A 4 Wireless LAN Tap Add Manually create a network profile to set ...
Page 282: ......
Page 318: ......
Page 322: ......
Page 323: ...P N 046 006959 07 1 0 ...