
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 640
P- 2.0.2 SYSTEM INITIALIZATION DOES BOT REACH 100% / I24 V-A
a) Spread the card cage side panels open then gently drop the front panel down to access the
motherbord’s white two (2) pin, +24V-A connector (Figure below).
Figure 128 – Card Cage TEST Connector (24V-A)
b)
Per the Figure above
, place the red meter lead into the right hand socket of the +24V-A connector.
ENSURE GOOD contact with the metal pin inside the socket!
c)
ENSURE the meter’s black lead REMAINS attached to chassis ground!
d)
If not already, turn the machine is ON (fan running)!
e) Between 23.0 and 28.0 volts DC AND STABLE i.e. does NOT change more than +/- 0.5 per minute?
Yes
Between 23.0 and 28.0 volts and stable! See procedure number
(page 641).
No IS NOT between 23.0 and 28.0 volts OR unstable! Proceed to
page 649
, procedure number
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...