2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 261
T- 1.8.2 VERIFY TEMERATURE CONTROL
This is NOT a routine Temperature calibration. Follow the procedure exactly to avoid error!
a) Install a Temperature meter (° C) in the dialysate lines and!
Figure right, flow through the meter MUST be
bottom to top!
b)
CLOSE THE SHUNT DOOR!
c) From the Service menu s
elect Calibrate Sensors → Temp Control.
d) ‘Sharply’ press ‘CONFIRM’ TWICE to turn the screen’s
[TEMP
DAC]
data box pale yellow / white.
e) Select
[TEMP DAC]
to turn it bright yellow
f) Set
[TEMP DAC]
to “180”.
g) ‘Sharply’ press ‘CONFIRM’.
[TEMP DAC]
MUST remain pale yellow/ white i.e. NOT gray!
h) ENSURING flow through the external indicator AND the Heater Switch is on, allow up to five (5) minutes
OR until the external meter reaches 35.5° C or more?
Yes
Meter 35.5° C or more! See procedure number
No Meter less than 35.5° C! With
[TEMP DAC]
pale yellow / white, see procedure number
T- 1.8.3 METER MORE THAN 35.5°C / VERIFY [TEMP DAC]
a) Allowing five (5) minutes after each
[TEMP DAC]
adjustment, continue the calibration until the external
meter is between 36.9 and 37.1° C.
b) Does
[TEMP DAC]
have to be set to 190 or more to between 36.9 and 37.1° C?
Yes
[TEMP DAC]
is 190 or more! See procedure number
No
[TEMP DAC]
less than 190! Proceed to
page 309
, procedure number
T- 1.8.4 METER LESS THAN 35.5° C / ISOLATE INCOMING WATER TEMPERATURE
If incoming water is less than 20.0° C the machine may not be able to maintain temperature.
a) Using a calibrated temperature meter measure INCOMING WATER from the RO!
b) More than 20.0° C?
Yes
More than 20° C! See procedure number
(page 262).
No
Less than 20.0° C! Incoming water temperature is too low! Once this problem is fixed, most
likely, temperature can be calibrated.
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...