2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 332
CO- 1.0.8.2 CONDUCTIVITY] MORE THAN 14.5 mS / ISOLATE DRAIN STREAM
.
a) ENSURE Loading Pressure (Acid Rinse Port gauge) continues to reach its high of more than 22 psi!
b)
IMPORTANT!
If a ‘Quick Connector’ (Figure
right) is present at the end of the drain tubing an
ADAPTOR is required!
c) As seen in the Figure (below right), point the drain
tubing opening
UP at 45°
holding it no higher
than two (2) feet above the floor!
IF POINTED
DOWN GRAVITY FLOW RESULTS IN
ERROR!
d) Watch for
ONE (1) FULL MINUTE
!
e) Consider BOTH scenarios below AND proceed
accordingly:
Scenario #1:
See Figure A. Approximately 30 ml pulses that stop between each and every
cycle i.e. Pulse → Stop → Pulse → Stop → etc.
Reconnect the drain then see
(page 332).
Figure A
Scenario #2:
See Figure B. If a continuous weak stream that LASTS for several seconds or
longer!
ENSURING the drain tubing opening was pointed up, proceed to
page 58
Figure B
CO- 1.0.8.3
DRAIN FLOW PULSE → STOP / [CONDUCTIVITY] MORE THAN 14.5 mS
A) Return BOTH concentrate connectors to their RINSE ports
B) Close the shunt door!
C)
IMPORTANT! Place the machine into RINSE!
D) Is the external flow indicator’s ‘bob’ moving at least ¼ up in the sight tube?
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...