2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 400
EOS- 4.0.0 PUMP SYMBOL REMAINS PINK / ISOLATE THE PUMP CIRCUITS
a)
IMPORTANT! Turn the machine OFF to prevent damage!
(page 398), return both pump cables to their distribution board positions.
c) See
EOS- 4.0.0, STEP #1
below:
EOS- 4.0.0, STEP #1
Isolate Actuator-Test Board
a)
IMPORTANT! Turn the machine OFF to prevent damage!
b) Swap in a known good Actuator-Test Board. To LOCATE the board refer to
(page 9).
c) If the problem was occurring in a Cleaning Program return to it. If not return to Dialysis Program.
d) Call debug screen 0. If debug does not appear press ’Escape’ then call screen 0.
e) Allow thirty (30) seconds THEN, WITHOUT LOOKING AWAY, watch for pump symbol for up two (2)
minutes. Does it remain white now?
Yes Remains white! The previous Actuator-Test Board is bad.
No Turns pink! Continue to
EOS- 4.0.0, STEP #2
(the previous Actuator-Test Board is okay).
EOS- 4.0.0, STEP #2 Isolate the Sensor Board cable:
a)
IMPORTANT! Turn the machine OFF to prevent damage
b) Swap in (or check
1
) a known good Sensor Board cable. To LOCATE the cable refer to
9).
1
To check the Sensor Board cable, which pump is issuing the EOS error?
IF ACID: NOTE
that one (1) ACID EOS connection (ONLY) will be checked and proceed to
page
SECTION 17- CHECKING THE SENSOR BOARD CABLE
. If the cable checks
good continue to procedure number EOS- 4.0.0, step #3.
IF BICARB: NOTE
that one (1) BICARB EOS connection (ONLY) will be checked and proceed to
page 524
SECTION 17 - CHECKING THE SENSOR BOARD CABLE
. If the cable
checks good continue to procedure number EOS- 4.0.0, step #3.
c) If the problem was occurring in a Cleaning Program return to it. If not return to Dialysis Program.
d) Call debug screen 0. If debug does not appear press ’Escape’ then call screen 0.
e) Allow thirty (30) seconds THEN, WITHOUT LOOKING AWAY, watch for pump symbol for up two (2)
minutes. Does it remain white now??
Yes
Remains white! The previous Sensor Board cable is bad
No Turns pink! Continue to
EOS- 4.0.0, STEP #3
next page.
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...