2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 34
F- 2.8.0 ISOLATE FLOW ERROR
a) Call debug screen 0 to see
Flow Error
. If EVER = 1, even if
just once, a Flow Error is present!
b) WITHOUT LOOKING AWAY, watch for two (2) minutes OR until
Flow Error
EVER = 1. TWO (2)
possible scenarios:
1)
IF (and ONLY if) Flow Error EVER = 1 even if just once:
A problem is indicated in the ‘bypass
circuit’! Proceed to
page 47
, procedure number
2)
IF Flow Error ALWAYS = 0 (NEVER = 1):
(page 34).
F- 2.8.1 FLOW ERROR ALWAYS = 0
a) Press and hold the “1” key for five
(5) seconds!
b)
Allow forty-five (45) seconds
BEFORE continuing to part
c!
c) Call debug screen 5.
d)
QdS
(Figure right)
REMAINS = 1 if
flow is stable. Watch it for one (1)
minute. TWO (2) possible
scenarios:
1)
If QdS is ALWAYS = 1!
Leaving the shunt door OPEN, see procedure number
(page 34).
2)
IF QdS is OR becomes = 0!
A problem is indicated in the ‘bypass’ circuit! Proceed to
page
, procedure number
F- 2.9.0 Qdl REMAINING BETWEEN 720 AND 880 / ISOLATE FOR A LEAKING BC VALVE
a) Call debug screen 4 and locate
PDIA
AND
ADIA
(Figure right).
b) BOTH should be between 2.0 and 8.5 AND remaining STABLE (i.e. does
NOT change more than 0.5 per minute). WITHOUT LOOKING AWAY,
watch for two (2) minutes! TWO (2) possible scenarios:
1)
IF (and ONLY if) PDIA AND ADIA
REMAINING
between 2.0 and 8.5 AND stable:
Proceed to
page 36
, procedure number
2) IF PDIA AND / OR ADIA IS NOT remaining between 2.0 and 8.5 AND / OR is NOT stable:
See
parts a THROUGH f next page:
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...