2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 264
Yes
Pin 6
is
9.0 volts OR MORE! See procedure number
(page 264).
No
Pin 6
LESS THAN 9.0 volts (bad signal from the Power Logic Board)! Proceed to
page 274
T- 2.0.1 PIN 6 MORE THAN 9.0 VOLTS
a) If in Heat Disinfect see part b. If in Dialysis Program ENSURE Dialysate Flow is on
*
!
*
If the Dialysate Flow on/off lamp is blinking flow is off. It MUST be on!
b) Figure below, at the top front edge of the Power Logic board, locate transistor
T5
.
Figure 44 – Power Logic Board / Transistor T5
c) Expecting almost 0 volts DC, measure at
T5’s TOP
solder point as shown above. Less than twenty
(20.0) volts DC
*
?
Yes
T5
LESS THAN 20.0 volts DC
*
(page 265).
No
T5
between 20.0 and 28.0 volts DC
*
! Proceed to
page 267
, procedure number
* Pay attention
to the meter’s
units!