
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 14
VOLTAGE DETECTOR CALIBRATION
A) Turn the machine OFF.
B) To prevent pulling cables loose, GENTLY open the card cage!
C) Behind the card cage, ENSURE the 24V POWER cable is securely plugged in!
D) Enter
Service Mode → Calibrate Monitor → Voltage Detection.
E)
Set your CALIBRATED volt meter to volts DC.
F)
Connect the meter’s black lead to chassis ground (refer to
.
G) Spread the card cage side panels open then gently drop the front panel down to access the
motherboard’s nine (9) pin TEST Connector (Figure below).
H) Measure from the TEST Connector’s +12V pin (pin 5, five pins from the left).
I) Select
the screen’s
[12 volt set]
data box. It turns bright yellow
J) Enter the MEASURED value in the box.
K) ‘Sharply’ press ‘CONFIRM’.
L) ENSURE the [
12 Volt Set]
data box is pale yellow /white.
M) Sharply press ‘CONFIRM’ again. Figure right, does an
“Operator Error” banner appear?
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...