
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 192
TROUBLESHOOTING A VALVE
A)
To avoid damage, turn the machine OFF!
B)
There is NO NEED to check all 17 valves!
Check ONLY the valve as
NOTED in the procedure that brought you here!!
C) See procedure number TSV- 1.0.0 below:
TSV- 1.0.0 CHECK DISTRIBUTION BOARD CONNECTIONS
a)
below
, remove the distribution board cover.
b) TWO (2) checks to ENSURE:
Check #1:
The ACTUATOR CABLE is plugged in SECURELY AND is not damaged!
Check #2:
The
NOTED
valve is plugged PROPERLY into its position.
If not, this most likely is the
problem!
Figure 34 –
Distribution Board
c) Before continuing to procedure number
(page 193) OR
(page 193) to LOCATE the
NOTED
valve.
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...