2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 462
TMP- 3.0.0 ISOLATE DIALYSATE PRESSURE TRANSDUCER
a) Figure below, trace the cable from distribution board position X9, “PDIAL” to the (blue) Dialysate
Pressure Transducer to ENSURE it has not been reversed connected with another hydraulic
component.
b) Call debug screen 1. ENSURE
FLWP
(upper right column) = 255 constant i.e. Dialysate Flow is “OFF”.
c) Remove the red dialyzer connector from the shunt door and place it at dialyzer level.
d) Call debug screen 10. Is
ACFS
(right corner) between 3.5 and 6.0?
Yes
ACFS
between 3.5 and 6.0! See procedure number
(page 463).
No
ACFS
is NOT between 3.5 and 6.0! Proceed to
page 464
, procedure number