2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 40
F- 3.0.0 ISOLATE VALVE #43 (FILLING PROGRAM) DRAIN FLOW
A) WITHOUT returning the Air Sensor’s connector yet, remove the resistor plug from the distribution board.
B) ENSURE debug screen 0’s Chamber #69 says “Air”.
C) Obtain a 1000 ml graduated cylinder!
D) Figure right, ENSURE a “Dial Valve Failure” OR “Act
Byp Valve Fail” banner NEVER appears!
E)
Call debug screen 1. FILACT MUST = 1 before continuing!
(page 40).
F- 3.1.0 FILACT = 1 / ISOLATE VALVE #43 FLOW
a) Figure right, if a ‘Quick Connector’ is present at
the end of the drain tubing an
adaptor
is
required!
b) Measure drain flow for two (2) minutes. 600 ml or
more collected?
Yes
600 ml or more! Reconnect the drain then proceed to
page 42
, procedure number
No Less than 600 ml! See procedure number
(page 41).