
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 418
FIL- 2.0.2 ‘BOB’ MOVING / ISOLATE POSSIBLE AIR LEAK (1)
Figure right, USING A FLASHLIGHT, look for air bubbles MOVING
through the dialyzer lines. Air seen?
Yes
Air seen! Proceed to
page 490
(page 418).
Figure 67 – Chamber #69 Air Check
FIL- 2.0.3 NO AIR IN DIALYZER LINES
Figure above
, USING A FLASHLIGHT, FOR TWO (2) MINUTES, watch for air bubbles MOVING into
Chamber #69
!
Air seen?
Yes
Air seen! Proceed to
page 490
SECTION 11 - INDUCED AIR LEAK TESTS
No air seen! See procedure number
(page 419).
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...