2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 373
CO- 8.9.0 ISOLATE POSSIBLE LEAKING HEAT EXCHANGER
a)
IMPORTANT!
Leaving the water on turn the machine OFF!
b) Per the Figure below, remove the front clear tubing from the Heat Exchanger’s nozzle.
Ignore a small
water crown that may appear!
c) Watch the open heat exchanger nozzle for up to three (3) minutes. Is water continually dripping,
possibly VERY slowly, from the Heat Exchanger’s nozzle?
Yes
Dripping from Heat Exchanger! TWO (2) procedures MUST be performed!
Procedure #1:
With the water turned OFF, REPLACE the Heat Exchanger;
Procedure #2:
See procedure
(page 551) to check incoming water pressure.
No
dripping
seen a)
Return the tubing to the Heat Exchanger.
b) From the Home screen, set the [Dialysate Flow] window to 500 ml/min and
press ‘CONFIRM’.
c) See (ABOVE) procedure number
(page 352).