2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 91
F- 8.3.0 BOTH ACFS VALUES ARE GOOD
a) Turn Dialysate Flow ON (Flow on/off lamp NOT blinking).
b) From the home screen, select the [Dialysate Flow] window.
c) Set [Dialysate Flow] to to 800 ml/min and press ‘CONFIRM’.
d) Allow one (1) minute BEFORE continuing to part e.
e) Call debug screen 10 to watch
ACFS
for one (1) minute. FOUR (4) possible scenarios 1) or 2) or 3) or
4) below:
1)
IF (and ONLY if) ACFS CYCLES between less than 0.5 to somewhere between 3.0 and
5.8, possibly every nine (9) seconds:
(page 91).
2)
IF (and ONLY if) ACFS REMAINS ALWAYS between 0 and 3.0:
Proceed to
page 104
3)
IF (and ONLY if) ACFS REMAINS ALWAYS between 3.0 and 5.8:
Proceed to
page 92
4)
IF ACFS REMAINS ALWAYS more than 5.8:
Proceed to
page 104
, procedure number
F- 8.4.0 ACFS CYCLING
Call debug screen 6. WITHOUT LOOKING AWAY, watch
BC Switch
(middle column)
for three (3) minutes. If it EVER becomes more than 400 OR less than 203, even if only
once, indicates a Flow Problem?
Yes
BC Switch
becomes more than 400 OR less than 203. Return to (ABOVE) procedure number
page 30
).
No
BC Switch
IS NEVER more than 400 OR less than 203. A flow problem is not presenting at this
time.
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...