2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 135
MOTORS- 4.5.0 ! EMPTY = 1 / ISOLATE MOTOR BRUSH CONNECTION
a) UNPLUG the NOTED motor from the distribution board!
b) Using a flashlight, check
inside
the motor’s distribution board position for ‘white’ corrosion or damaged
male
pins.
Damage indicates the distribution board may need to be replaced!
c) Figure right, remove the plastic cap from the female
distribution connector.
•
If the
NOTED
motor is the
Deaeration Motor
it
MUST be connected between
pins 1 and 3 i.e. top
and middle
•
If the
NOTED
motor is the
Flow Motor
,
it MUST be
connected between
pins 1 and 5 i.e. top and bottom.
d) Set your CALIBRATED voltmeter to
RESISTANCE
(Ώ)
.
e) Touch the meter’s leads together. The meter MUST read less than 0.4
Ω
! Subtract this from the
subsequent measurement.
f) Measure,
INSIDE
the female connector, BETWEEN the soldered terminals. Less than ten (10)
Ω
?
Yes
Less than 10 Ω! See procedure number
(page 136).
No
More than 10 Ω! Bad brushes or motor.