
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 69
F- 7.4.44 PART 1 MORE THAN 800 ML / PART 2 BAD FLOW / RESTRICTED VALVE #37 OR #33
Balancing Chamber Valve #37 and/or Valve #33 is not opening. Also, the Actuator Cable or distribution
board may be bad. These procedures isolate between these components:
A) Turn the machine OFF!
B)
Per the Figure below
, replace
Valve #37
with a known good valve.
C) Plug the new Valve #37’s connector into Valve #30’s distribution board position, “V30”.
D) Turn the machine on but
DO NOT
press any other keys!
E) With “Select Program” up, measure flow, from the RED connector, into the cylinder for one (1) minute.
More than 800 ml collected?
Yes
More than 800 ml collected! Problem solved! The previous Valve #37 is bad.
No
Less than 800 ml collected! See procedure number
(page 69).
.
F- 7.4.45 LESS THAN 800 ML COLLECTED / ISOLATE VALVE #33
A) Turn the machine OFF!
B) Plug the new valve #37’s connector into its distribution board position, “V37”!
C)
Per the Figure above
, replace
Valve #33
with a known good valve.
D) Plug the new Valve #33 into Valve #30’s distribution board position, “V30”!
Parts E and F next page
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...