2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 378
CO- 10.0.00 TROUBLESHOOTING ONLINE CLEARANCE (OLC)
OLC estimates UREA clearance (Kt/V) by measuring Conductivity pre and post dialyzer:
a) Return the dialyzer connectors to the shunt but
LEAVE THE DOOR OPEN till
instructed!
b)
Per the Figure below
, trace the wires from distribution board positions #13 (Post Dialyzer Cond
Cell #13) AND #44 (Post Temperature Sensor NTC #44), to the sensors to perform TWO (2) checks:
CHECK #1:
Check the entire length for insulation damage.
Damage will cause OLC problems and,
if damaged, the sensor should be replaced!
CHECK #2:
ENSURE the connectors ARE NOT reverse connected another component! ’!
c) From the Home screen, set [Dialysate Flow] to 500 ml/min and press ‘CONFIRM’.
Figure 62 – Conductivity / OLC Components