2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 30
F- 2.0.0 ISOLATE VALVE ERROR / BALANCING CHAMBER VALVES CYLING
a) Call debug screen 1 to ENSURE
FILACT
REMAINS = 0!
b) Call debug screen 0.
c) Figure right, watch the Balancing Chamber valve
‘dots’ for
thirty (30) seconds
or until if they
CYCLE between white and blue.
d) TWO (2) possible scenarios 1) or 2 below:
1)
IF (and ONLY if) the ‘dot’s cycle, possibly every
nine (9) seconds:
Proceed to
page 32
, procedure
2) IF the ‘dots’ NEVER cycle i.e. four (4) STAYING blue AND four (4) STAYING white OR all eight
(8) STAYING white:
(page 30).
F- 2.1.0 THE BALANCING CHAMBER (BC) ‘DOTS’ ARE NOT CYCLING
Call debug screen 1. Is
VERR
(right column, bottom) = 0?
Yes
VERR
(page 30).
No
VERR
= 1 OR more!
See parts a AND b below:
a) Call debug screen 0 (Figure right) and locate
Valve Error
(
2
nd
window down). IGNORE the top
Flow Error
window.
b) WITHOUT LOOKING AWAY, ignoring a ‘blink’ to 1 that lasts less than (2) two seconds,
watch
Valve Error
for one (1) minute or until if it = 1 for LONGER THAN two (2) seconds.
TWO (2) possible scenarios:
1)
IF (and ONLY if) Valve Error = 0 OR ‘blinks’ to 1 for less than two (2) seconds:
Proceed to
page 195
TROUBLESHOOTING VALVE ERRORS IN DIALYSIS
2)
IF Valve Error EVER = 1 LONGER THAN two (2) seconds!
Proceed to
page 652,
F- 2.2.0 BC VALVES NOT CYCLING / VERR = 0
Call debug screen 2. Is
! EMPTY
(left column) = 1?
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...