2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 24
F- 1.0.4 FILACT = 1 OR CHAMBER #69 SAYS “AIR”
a)
FIGURE BELOW
, place one of the plugs, from the
FOUR-RESISTOR SET
, into the Air
Sensor’s distribution board
position i.e.
5
th
position
*
from the LEFT.
*
NOTE:
If CBE modified the resistor plugs into the CBE board pin for pin!
b) Based on Chamber #69’s text box now. TWO (2) possible scenarios:
1) IF (and ONLY if) it says “No Air” always:
See
(page 26).
2) IF says “Air” OR toggles between “Air” and “No Air”:
To avoid unnecessary work, perform
parts a THROUGH c below:
a)
BE CERTAIN
the resistor plug is placed PROPERLY in the
5
th
position
from the left!
b) If CBE modified, using a flashlight, ENSURE the resistor plugs into the CBE board, pin for pin.
ENSURE the CBE board’s TOP pin is covered by the resistor plug!
c) If (and ONLY if) the text box now says “No Air” always see
26). If it EVER says “Air” see
part d next page
!
Part d next page