2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 338
c)
IMPORTANT! Place the machine into RINSE!
d) Is the external flow indicator’s ‘bob’ rising at least ¼ way up in the sight tube?
Yes
‘Bob’ moving! Proceed to
page 353
, procedure number
No ‘Bob’ NOT moving! See parts a AND b below:
a) Return to Dialysis Program
(“Select Program” → ‘Dialysis’ → ‘CONFIRM’)!
b) Proceed to
page 20
SECTION 1 – FLOW ERRORS IN DIALYSIS