2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 42
F- 3.2.0 SIX HUNDRED (600) ML OR MORE IN FILLING PROGRAM
a) Return the Air Sensor’s female connector to
the
5
th
distribution board position from the
left.
b) Call debug screen 0.
If the Air Sensor is
plugged in properly, Chamber #69’s
box says “No Air”.
(page 42).
F- 3.2.1 CHAMBER #69 = “No Air”
a) This procedure uses a psi gauge. ENSURE it reads 0 psi before installing it.
b) Return BOTH concentrate connectors to their RINSE ports!
c) Figure right, tee the gauge between the Flow Pump's
OUTPUT nozzle and its WHITE tubing.
d)
Tie wrap both sides of the gauge tubing to
prevent leaks and false readings!
e)
(page 42).
F- 3.2.2 ISOLATE FLOW PUMP
a)
Place the machine in RINSE!
b) Watch for one (1) minute to ENSURE a “No Water” alarm NEVER occurs!
c) Call debug screen 0. If debug does not appear press ‘Escape’ then call screen 0.
d) Allow Valve #43’s ‘dot’ (Figure right) to turn blue then WHITE again! While
white, does pressure CYCLE, about every three (3) seconds, to between 35
and 36 psi?
Yes
Between 35 and 36 psi! See procedure number
(page 43).
No
NOT between 35 and 36 psi! ENSURING the machine was in RINSE AND no leaks, TWO
(2) possible scenarios below:
1)
IF (and ONLY if) pressure is too low:
DO NOT calibrate instead proceed to
page
, procedure number
3.
2) IF pressure is too high:
Perform parts a AND b below:
a) Per the Figure (above), adjust Valve #78 until pressure cycles to between 35 and 36
psi!
(page 43).
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...