2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 445
TMP- 1.0.5 LOADING PRESSURE OKAY / ISOLATE AIR REMOVAL SYSTEM
a)
Turn the water OFF!
b) The Deaeration gauge is used next.
ENSURE
it reads 0
inHg before installing it!
c) Allow a “No Water” alarm to occur THEN, after one (1)
minute, the deaeration motor stops.
d) Figure right, tee the gauge into the Inlet (clear tubing)
side of the Deaeration Pump.
e)
IMPORTANT! Turn the water on and allow
the “No Water” alarm to go away!
f) Is Deaeration Pressure OKAY? Refer to
(page 667) for what pressure should be.
Yes
Deaeration pressure OKAY! See procedure number
(page 446).
No Deaeration Pressure in NOT OKAY! ENSURING a “No Water” alarm is NOT presenting,
NOTE this page number, as you will return here, THEN proceed to
page 498